Water level Logger

VLL-999 Recording Transmitter is able to be widely used in petroleum, chemi-industry, power station, water conservancy, city flood control and city water supply and drainage, etc. It is supplied by battery, could work without manual control. The transmitter could record level or pressure change continuously. By setting proper data collection interval, the transmitter could record from several days to several years. Very low power consumption is another advantages, one AA baterry 3.6V/2400mAh could work many years continuously. VLL-999 has an absolute pressure sensor inside, to measure lower level, please be careful for atmosphere change. We suggest to install a barameter or another VLL-999 Recording Transmitter to record air pressure. And then calculate two pressure difference of two instruments by computer, that is the final level value; Please pay attention that get VLL-999 back from the measurement point, then remove end cap and read record through data interface.

Water level Logger

VLL-999 Recording Transmitter is able to be widely used in petroleum, chemi-industry, power station, water conservancy, city flood control and city water supply and drainage, etc. It is supplied by battery, could work without manual control. The transmitter could record level or pressure change continuously. By setting proper data collection interval, the transmitter could record from several days to several years. Very low power consumption is another advantages, one AA baterry 3.6V/2400mAh could work many years continuously. VLL-999 has an absolute pressure sensor inside, to measure lower level, please be careful for atmosphere change. We suggest to install a barameter or another VLL-999 Recording Transmitter to record air pressure. And then calculate two pressure difference of two instruments by computer, that is the final level value; Please pay attention that get VLL-999 back from the measurement point, then remove end cap and read record through data interface.

Features:

Features:

  • The latest microprocessing technique; better accuracy and resolution for pressure and temperature signals; digital compensation for both linearity and temperature error of pressure sensor;
  • Good data security due to nonvolatilememory units;
  • Single 3.6V/2400mAh Lithium cell for power supply; low current s30uA for Standby Mode; long-term collection available;
  • Full stainless steel housing; multi-track O-ring sealing; easy to change battery and read data;
  • Complete recording to pressure, temperature and the happening time;
  • Flexible recording time setting, the min. ls, the max. 255 hours 59 minutes 59 seconds;
  • The max. 50,000 Recording 
  • Bus line RS485 protocol; freely wake up transmitter to set and read pressure value, temperature value and recording;
  • Factory setting restore available, could reduce cost when the user makes incorrect setting or maintenance;
  • High stable design, use hardware watch-dog and multi-track software inspection to reduce system halted;
  • Protection class IP68, long-term submersible application available; 
  • High accuracy class at wider temperature range - 30°C ~80°C

Specifications

Specifications

Model  VLL-999

Pressure range 

0-10m H2O / 0-20m H2O / 0-30m H2O / 0-100m H2O

Accuracy

0.5%

Overpressure 

2xpressure range 

Power supply 

Cylinder 3.6V/2400mAh Lithium cell 

Output signal 

RS485 digital 

Environment temp. 

-30°C ~-80°C·

Error1 

10°C~40°C ±0.10%FS(typ.) ±0.2%FS(max.)
30°C ~80°C ±0.15%FS(typ.) ±0.3%FS(max.)

Stability 

<±0.2%FS/year (typ.) 

Resolution 

0.01%FS 

Temperature resolution 

0.1°C (-30°C ~80°C) 

Temperature accuracy 

±1°C

Record interval 

1s~255 hours 59 minutes 59 seconds 
(1s~10 days 15 hours 59 minutes 59 seconds) 

Record data quantity 

Recording stops when the records are 50,000 (including date, time,temperature and pressure) 

Static current 

(no collection, no recording, no communication)≤30uA 

Battery life 

Typ. (once per hour)about 10 years 

Construction material 

Housing: SS 316L   O-ring: Viton 

Protection class 

IP68 

Others 

Weight: about 500g Shock:20g 20Hz~5000Hz
Resolution: 100MO@50V Impact: 20g llms  

1. Error calculation: the absolute value of max. error at working temperature and pressure.

Accessories

Accessories

Data Transfer Cable

We can transfer the Data through Data  Transfer Cable